Proceedings of the 10th Convention of the European Acoustics Association Forum Acusticum 2023 Politecnico di Torino Torino, Italy September 11 - 15, 2023 |
|
Abstract Objective: Evaluate the effectiveness of two design approaches in a newly developed smartwatch-based alarm system: (1) using audiovisual spatial colocalization and (2) adding haptic information. Background: The volume, design, and pervasiveness of existing ICU alarms can make it difficult for clinicians to quickly distinguish alarms’ importance and meaning. Method: We compared the performance of 30 participants using newly developed ICU smartwatch alarms with two sensory modalities (visual and audio) against alarms with three sensory modalities (adding haptic cues). Additionally, we compared participant performance in two implementations of the audio modality: colocalized with the visual cue on the smartwatch’s low-quality speaker versus delivered from a higher quality speaker located two feet away from participants. Results: Participants were 13.5% (0.35s) faster at responding to alarms when auditory information was delivered from the smartwatch instead of the higher quality external speaker. Meanwhile, adding haptic information to alarms improved response times to alarms by 14.7% (0.31s) and response times on their primary task by 9.5% (.08s). Participants also rated learnability and ease of use higher for alarms that included haptics. Conclusion: Audiovisual colocalization and multisensory alarm design improve user response times without compromising accuracy, while also improving user performance in concurrent cognitively demanding tasks. |
||||||||||||||||||||