Proceedings of the 11th Convention of the European Acoustics Association Forum Acusticum / EuroNoise 2025 Málaga, Spain June 23 - 26, 2025 |
|
Abstract Acoustic holograms are used in transcranial ultrasound to tailor the acoustic wavefront generated by a singleelement transducer compensating skull aberrations and focusing the beam over brain structures. However, current hologram design relies on full-wave simulations where acoustic properties are known by computed tomography (CT). Since exposing patients to ionizing radiation is undesirable, we propose the use of zero echo time (ZTE) magnetic resonance images (MRI) to construct the patient-specific lens. This MRI sequence enables the visualization of tissues with short T2 values, such as bone. By establishing a linear correlation between CT Hounsfield Units (HU) and ZTE intensity values, pseudo-CT images are generated. Furthermore, transducer positioning has been optimised. Temporal bone window is established as a suitable surface for propagating acoustic waves, due to its reduced thickness and flat morphology. Results show that peak pressure at the focus is reduced by 5.38% when using pseudo-CT holograms respect to CT-holograms. The obtained treated volumes considering acoustic properties of the pseudo-CT and those of the CT show a difference of 0.18%. This approach can be applied in holographic wearable devices for transcranial ultrasound therapies such as blood-brain barrier opening or neuromodulation to avoid the use of ionising radiation during the therapeutic treatment. |
||||||||||||||||||||