Overpressure Dependence of Sub-Harmonic Generation from Contrast Agent SonoVue Microbubbles
Jun Wu
Tingbo Fan
Di Xu
Dong Zhang
Abstract
Sub-harmonic component generated from microbubbles has been proven to be potentially used in noninvasive blood pressure measurement. In the present work, the overpressure dependence of the sub-harmonic generation from a commercial ultrasound contrast agent SonoVue was theoretically
and experimentally investigated with varied excitation pressure amplitudes and pulse lengths. With 4-MHz ultrasound excitation at an applied acoustic pressure amplitude of 0.2 MPa, the measured sub-harmonic amplitude exhibited a decreasing change as the overpressure increases; while non-monotonic
change was observed for the applied acoustic pressures of 0.3 and 0.4 MPa, and the peak position in the curve of the sub-harmonic response versus the overpressure was shifted towards higher overpressure as the excitation pressure amplitude increases. Furthermore, the exciting pulse with long
duration could lead to better sensitivity of the sub-harmonic response to the overpressure. The measured results were interpreted by the numerical simulations based on the Marmottant model. This work might provide a solid proof for the optimization of the noninvasive blood pressure measurement
using sub-harmonic generation from microbubbles.