Inter Laboratory Comparison of Environmental Noise Measurement
Jurij Prezelj
Mirko ÄŒudina
Abstract
Results of an inter laboratory comparison for environmental noise measurements are presented in this article. 18 laboratories took part in this study. A purpose of individual measurement was to quantify environmental noise levels, calculate noise descriptors and compare them with community
noise regulations. Measurements and calculations were performed according to standards ISO 1996-1:2005, ISO 1996-2:2007 and national legislation. A loudspeaker was used as a reference noise source. It was mounted at the top of the hayloft just below the roof. The loudspeaker emitted a pink
noise, representing a constant broadband noise source. The noise source was operating from 8:00 in the morning to 16:00 at the afternoon, each day for 4 weeks. During this time, laboratories performed all required measurements. Each laboratory performed measurements independently from other
laboratories and according to its own procedures. Measurement results of equivalent noise level LAeq, percentile values L1, and calculation of noise descriptors Ld and Lden were compared. An estimation of measurement uncertainty
for a given environmental conditions and measurement setup conditions is given.