Issue
Acta Acustica Volume 99, Issue Number 4, 2013
Number of page(s)8
DOIhttps://doi.org/10.3813/AAA.918637
 
 
 
Ear Canal Properties of Children: Dimensions of Ear Canals and Simulation of the Input-Impedance
 
 
  • Janina Fels
  • Jan Paprotny
 
 
Abstract
Today, only a few standardized couplers (i.e. a 2 ccm coupler (IEC 60126)) are used to fit hearing aids for infants, children, and adults. If we want to develop couplers that improve the fitting process for different age groups, we have to study the ear canal properties of infants and children in detail. The smaller ear canal of children leads to different input impedances than those of adults. In this study CAD-models were generated using CT-scans (computed tomography) of young children and adults. Using the CAD-models the ear canal properties and geometry are studied. A simulation of the ear canal impedance using FEM (Finite-Element-Method) was performed and analyzed in this study.